FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE

MDR report key: 1823915 · Received January 16, 2008

Report

Report Number
1527736-2008-00313
Event Type
Malfunction
Date Received
January 16, 2008
Date of Event
August 24, 2007
Report Date
November 20, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH AN SUFFICIENT H-CRIMP, WITH THE FIRING MECHANISMS DAMAGED AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. THE RETURNED CARTRIDGE WAS LOADED INTO A TEST DEVICE AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON HOW THE DEVICE GOT DAMAGED WITH THE INFO PROVIDED, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE LOCKED AND DID NOT FIRE. COMPLETED WITH THE SAME LIKE PRODUCT. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE GDW ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA C4EH7L

Patients

Seq Age Sex Outcome Treatment
1