ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE
Report
- Report Number
- 1527736-2008-00313
- Event Type
- Malfunction
- Date Received
- January 16, 2008
- Date of Event
- August 24, 2007
- Report Date
- November 20, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4): EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH AN SUFFICIENT H-CRIMP, WITH THE FIRING MECHANISMS DAMAGED AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. THE RETURNED CARTRIDGE WAS LOADED INTO A TEST DEVICE AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON HOW THE DEVICE GOT DAMAGED WITH THE INFO PROVIDED, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT THE DEVICE LOCKED AND DID NOT FIRE. COMPLETED WITH THE SAME LIKE PRODUCT. THERE WAS NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE | GDW | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | C4EH7L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |