FDA Adverse Event Malfunction Summary report: N

HARMONIC SCALPEL GENERATOR

MDR report key: 1823841 · Received December 12, 2007

Report

Report Number
1527736-2007-08465
Event Type
Malfunction
Date Received
December 12, 2007
Date of Event
November 13, 2007
Report Date
November 14, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE ANALYSIS SITE CONFIRMED THE CUSTOMER COMPLAINT AND TO CORRECT THE COMPLAINT THE MAIN PC BOARD WAS REPLACED. THE DEVICE HISTORY RECORD HAS BEEN REVIEWED VIA THE DATABASE. THE UNIT HAS PASSED ALL QA INSPECTIONS. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO AN UNKNOWN PROCEDURE THE GENERATOR EMITTED A SOLID TONE DURING THE USE. NO PATIENT CONSEQUENCES WERE REPORTED. DEVICE RETURNED TO THE INTERNATIONAL SERVICE CENTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC SCALPEL GENERATOR LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE