OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2010-01263
- Event Type
- Other
- Date Received
- August 27, 2010
- Date of Event
- August 4, 2010
- Report Date
- August 4, 2010
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE POD WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS SEEN IN THE CANNULA. THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE OMNIPOD USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES. THE PATIENT REMEDIED THE SITUATION BY REMOVING AND REPLACING THE DEVICE PER USER INSTRUCTIONS.
THE CUSTOMER REPORTED "HIGHER THAN EXPECTED BG LEVELS WITHOUT INDICATION OF ALARM". OVER AN EIGHT HOUR PERIOD, HER BG'S ROSE SIGNIFICANTLY (FROM 325-575MG/DL). SHE, THEREFORE, "DOESN'T THINK THE POD IS DELIVERING INSULIN ACCURATELY". THE CANNULA WAS REPORTEDLY KINKED, THOUGH THE POD DID NOT INITIATE AN ALARM. THE POD WILL BE RETURNED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30346 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Other |