ACCU-CHEK ® AVIVA TEST STRIPS
Report
- Report Number
- 1823260-2010-05211
- Event Type
- Malfunction
- Date Received
- September 2, 2010
- Date of Event
- August 16, 2010
- Report Date
- September 30, 2010
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- LFR
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ES
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE EVENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS FOR THE SUSPECT DEVICE USED IN SYSTEM 2 (LOT NUMBER 202699, EXPIRATION DATE 10/31/2011). (B)(4).
THE ACCOUNT STATED THAT THE ARCHITECT C16000 ANALYZER GENERATED A CALCIUM RESULT OF <2 MG/DL. THE RESULT WAS FLAGGED AND WAS NOT REPORTED OUT OF THE LAB. THE SAMPLE WAS REPEATED AND A RESULT OF 9.9 MG/DL WAS GENERATED. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT. NO FURTHER PATIENT INFORMATION WAS AVAILABLE.
CALLER STATES NEONATE PATIENT RECEIVED RESULTS OF 36 MG/DL ON AVIVA SYSTEM 1 AND 37 MG/DL ON AVIVA SYSTEM 2 WHILE A COMPARISON LAB RETURNED AS 60 MG/DL. THE REPORTED VALUES WERE OBTAINED WITHIN 10 MINUTES OF EACH OTHER. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK ® AVIVA TEST STRIPS | BLOOD GLUCOSE MONITORING TEST STRIPS | LFR | ROCHE DIAGNOSTICS | NA | 202699 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 002 DA |