Summary report: N
INOMAX DS (DELIVERY SYSTEM)
MDR report key: 1823749
·
Received August 26, 2010
Report
- Report Number
- 3004531588-2010-00050
- Event Type
- Malfunction
- Date Received
- August 26, 2010
- Date of Event
- July 28, 2010
- Report Date
- August 26, 2010
- Manufacturer
- INO THERAPEUTICS LLC/IKARIA
- Product Code
- MRN
- PMA / PMN Number
- K061901
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ON (B)(6) 2010, A RESPIRATORY THERAPIST REPORTED "DEVICE WAS MAKING A HISSING NOISE" FROM INOMAX DS (B)(4). EVALUATION SUMMARY: THE INVESTIGATION OF THE DEVICE IS COMPLETE AND IS AS FOLLOWS: THE DEVICE WAS CONFIRMED TO HAVE A LEAK THAT WAS TRACED TO THE PRESSURE SWITCH. THE PRESSURE SWITCH WAS REPLACED AND IT WAS CONFIRMED THE LEAK STOPPED AND WAS CORRECTED. THE ROOT CAUSE OF THE INCIDENT WAS A FAILED PRESSURE SWITCH.
Description of Event or Problem · 1
ON (B)(6) 2010, A RESPIRATORY THERAPIST REPORTED THE INOMAX DS (B)(4) DEVICE WAS MAKING A HISSING NOISE. THE DEVICE WAS NOT ON A PATIENT AND NO ADVERSE EVENT WAS REPORTED BY THE RESPIRATORY THERAPIST. THE DEVICE WAS REMOVED FROM SERVICE BY THE CUSTOMER AND RETURNED TO THE COMPANY FOR INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INOMAX DS (DELIVERY SYSTEM) | APPARATUS, NITRIC OXIDE DELIVERY | MRN | INO THERAPEUTICS LLC/IKARIA | 10003 |