NEEDLE ECLIPSE 30X1/2
Report
- Report Number
- 2243072-2023-02116
- Event Type
- Malfunction
- Date Received
- November 30, 2023
- Date of Event
- November 3, 2023
- Report Date
- January 18, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- FMI
- UDI-DI
- 30382903057574
- PMA / PMN Number
- K010188
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4): INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. PATIENT PROBLEM CODE: F26 ¿ NO HEALTH CONSEQUENCES OR IMPACT. DEVICE PROBLEM CODE: A140901 - COMPLETE BLOCKAGE.
BASED ON DEVICE HISTORY RECORD REVIEW, NO ABNORMALITY WAS OBSERVED DURING THE PRODUCTION OF THE AFFECTED BATCHES. CURRENT CONTROL THERE IS A VISUAL INSPECTION ON CLOGGED NEEDLE AT THE QA OUTGOING INSPECTION PER MQA-052/A AND VISUAL INSPECTION ON CLOGGED NEEDLE AT THE ASSEMBLY IN-PROCESS INSPECTION PER MFG-043/I. ACTUAL ROOT CAUSE COULD NOT BE DETERMINED AS NO SAMPLE WERE RECEIVED FOR INVESTIGATION. THE COMPLAINT WILL BE RE-OPENED AND RE-INVESTIGATED WHEN SAMPLE IS RECEIVED. H3 OTHER TEXT : SEE H10 MANUFACTURE NARRATIVE.
MATERIALS#: 305757 BATCH#: 2341255. IT WAS REPORTED BY THE CUSTOMER THAT INSTANT BLOCKAGE WHEN TRYING TO ADMINISTER MEDICINE THROUGH NEEDLE. VERBATIM: INSTANT BLOCKAGE WHEN TRYING TO ADMINISTER MEDICINE THROUGH NEEDLE.
MATERIALS#: (B)(4) BATCH#: 2341255 IT WAS REPORTED BY THE CUSTOMER THAT INSTANT BLOCKAGE WHEN TRYING TO ADMINISTER MEDICINE THROUGH NEEDLE. VERBATIM: INSTANT BLOCKAGE WHEN TRYING TO ADMINISTER MEDICINE THROUGH NEEDLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1940730 | NEEDLE ECLIPSE 30X1/2 | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | BECTON DICKINSON | 2341255 | 30382903057574 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |