Narratives
CONTINUATION OF D10: PRODUCT ID 3708220 LOT# SERIAL# (B)(6). IMPLANTED: (B)(6)2009.EXPLANTED: PRODUCT TYPE EXTENSION PRODUCT ID 3776-60 LOT# SERIAL# (B)(6). IMPLANTED: (B)(6) 2009. EXPLANTED: PRODUCT TYPE LEAD PRODUCT ID 3487A-45 LOT# V101634 SERIAL# IMPLANTED: (B)(6) 2009. EXPLANTED: PRODUCT TYPE LEAD PRODUCT ID 3487A-45 LOT# V183264 SERIAL# IMPLANTED: (B)(6) 2009. EXPLANTED: PRODUCT TYPE LEAD SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 3708220, SERIAL/LOT #: (B)(6), UBD: 06-MAY-2013, UDI#: (B)(4). PRODUCT ID: 3776-60, SERIAL/LOT #: (B)(6). UBD: 19-SEP-2011, UDI#: (B)(4) ; PRODUCT ID: 3487A-45, SERIAL/LOT #: (B)(6), UBD: 31-MAR-2012, UDI#:(B)(4) ; PRODUCT ID: 3487A-45, SERIAL/LOT #: (B)(6), UBD: 01-DEC-2012, UDI#: (B)(4). THIS REGULATORY REPORT IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW AS PART OF REMEDIATION PLAN 411. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCES ON THE DAY OF SURGERY. IMPEDANCES READING 17,000-40,000 ON LEAD SEATED IN 0-7. UNABLE TO DETERMINE WHICH LEAD IS IMPACTED AT THIS TIME DUE TO NO FLUOROSCOPY TAKING PLACE DURING BATTERY REVISION. MULTIPLE IMPEDANCE CHECKS AND CONNECTIVITY CHECK WHICH RESULTED IN ONE CONTACT SHOWING POOR CONNECTION. MD ATTEMPTED TO RESEAT LEAD MULTIPLE TIMES PRIOR TO WRENCHING IT IN. THE ISSUE IS ONGOING. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE (REP). IT WAS REPORTED THAT REP WAS ABLE TO ESTABLISH PROGRAMMING THAT COVERED THE PATIENT¿S CHRONIC PAIN AREAS AT POST-OP APPOINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1939714 | INTELLIS | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97715 | 00763000315467 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female | "SEE H10...." |