FDA Adverse Event
Malfunction
Summary report: N
SERIES 20000 LEGACY
MDR report key: 1823684
·
Received August 27, 2010
Report
- Report Number
- 2028159-2010-01620
- Event Type
- Malfunction
- Date Received
- August 27, 2010
- Date of Event
- July 30, 2010
- Report Date
- July 30, 2010
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K911808
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY SERVICE REP EXAMINED THE SYSTEM AND COULD NOT DUPLICATE THE PROBLEM REPORTED. THE POWER SUPPLY WAS REPLACED AS A PREVENTIVE MEASURE. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THE POWER SUPPLY HAS BEEN RECEIVED AND IN-HOUSE TESTING IS IN PROGRESS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADDITIONAL REPORTABLE INFO BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
ADVERSE EVENT(S): "NO PT INJURY REPORTED" (NO CONSEQUENCE OR IMPACT TO PT). PRODUCT PROBLEM(S): "SYSTEM SHUT DOWN" (LOSS OF POWER). THE NURSE REPORTED THE SYSTEM SHUT DOWN DURING SURGERY. THE SYSTEM WAS REBOOTED AND THE CASE WAS COMPLETED AS PLANNED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SERIES 20000 LEGACY | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | STTL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |