FDA Adverse Event Malfunction Summary report: N

HS HAND CONTROLLED CURVED SHEA

MDR report key: 1823647 · Received November 16, 2007

Report

Report Number
1527736-2007-07782
Event Type
Malfunction
Date Received
November 16, 2007
Report Date
October 8, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: TWO DEVICES (A-B) WERE RETURNED FOR ANALYSIS. BOTH DEVICES WERE RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS. DEVICE B ADDITIONAL INFORMATION: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICES DID NOT WORK FROM THE BEGINNING. THE ERROR CODE "SHEARS DEFECT" APPEARED AT THE SCREEN. NO PATIENT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS HAND CONTROLLED CURVED SHEA GEI ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA D4HF0W

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR