FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® CAT PLUS BLOOD COLLECTION TUBES

MDR report key: 18234708 · Received November 29, 2023

Report

Report Number
9617032-2023-01703
Event Type
Malfunction
Date Received
November 29, 2023
Date of Event
November 2, 2023
Report Date
November 7, 2023
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
30382903678373
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR CAP COLOR VARIATION WITH THE INCIDENT LOT WAS NOT OBSERVED. NO BD PRODUCT WAS DISPLAYED IN THE PHOTO. ADDITIONALLY, 100 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO CAP COLOR VARIATION AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CAP COLOR VARIATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® CAT PLUS BLOOD COLLECTION TUBES, THE SPECTROMETER USED TO PRESCREEN TUBE COLOR COULDN¿T IDENTIFY CORRECT COLOR. AT BARNET GENERAL HOSPITAL WE FOUND TWO DISTINCT DIFFERENCES IN COLOR OF CAP WHICH REGISTERED A 20 POINT DIFFERENCE ON SPECTROMETER. LIGHTER (OLD) LOT # 2122533 DARKER (NEW) LOT # 3052236.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2137105 BD VACUTAINER® CAT PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3052236 30382903678373

Patients

Seq Age Sex Outcome Treatment
1 Unknown