FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE

MDR report key: 18234632 · Received November 29, 2023

Report

Report Number
1213809-2023-01364
Event Type
Malfunction
Date Received
November 29, 2023
Date of Event
November 6, 2023
Report Date
February 7, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMI
UDI-DI
30382903059165
PMA / PMN Number
K951254
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR 9213832 - FOLLOW UP MDR FOR DEVICE EVALUATION. DURING REVIEW OF THE PROCESS, PROCESS VARIATIONS DURING THE NEEDLE LUBRICANT APPLICATION CAN CREATE CLOGGED NEEDLES. IT COULD BE POSSIBLE SOME SAMPLES ARE ESCAPES FROM THE INCIDENT THAT OCCURRED DURING PRODUCTION. CORRECTIVE AND PREVENTATIVE ACTIONS HAVE BEEN IMPLEMENTED. SEVERAL QUALITY INITIATIVES HAVE BEEN IMPLEMENTED ON OUR MANUFACTURING LINE TO ENSURE THAT THE NEEDLE LUBRICANT APPLICATION IS PROPERLY APPLIED DURING THE MANUFACTURING PROCESS. ADDITIONALLY, A MONITORING PROGRAM IS ALSO IN PLACE TO VERIFY THE NEEDLE LUBRICANT IS APPLIED UNIFORMLY TO THE NEEDLE. WE APPRECIATE YOU TAKING THE TIME TO BRING THIS OBSERVATION TO OUR ATTENTION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLES ARE CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER THAT CANNOT GET ANY AIR TO PRESS THROUGH THE SYRINGE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1941561 BD SAFETYGLIDE SAFETYGLIDE NEEDLES FMI BECTON DICKINSON MEDICAL SYSTEMS 2025239 30382903059165

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown