HARMONIC ACE 36CM W ERG HANDLE
Report
- Report Number
- 3005075853-2010-05030
- Event Type
- Malfunction
- Date Received
- September 2, 2010
- Report Date
- August 27, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE A AND C WERE RETURNED IN GOOD PHYSICAL CONDITION. THE DEVICES WERE TESTED WITH A GENERATOR AND BOTH WERE FUNCTIONAL. DURING INSPECTION, THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. THE DEVICE (B) WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DURING FUNCTIONAL TESTING, AN ERROR CODE 5 WAS DISPLAYED . THE DEVICE WILL STOP ACTIVATE (AND THEREFORE NOT CUT AND COAGULATE), AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A PROBABLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. DEVICE A BATCH G9JL1L. MFG DATE 3-9-10. EXP DATE 2-9-15. DEVICE B BATCH F9J420. MFG DATE 12-15-09. EXP DATE 11-15-14. DEVICE C BATCH G9J88W. MFG DATE 1-22-10. EXP DATE 12-22-14.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME ADDITIONAL INFORMATION REQUESTED: (B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPY PROCEDURE, THE SCISSORS WAS NOT COAGULATING. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36CM W ERG HANDLE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | 4RP57 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |