FDA Adverse Event Malfunction Summary report: N

CENTURY BED

MDR report key: 1823341 · Received August 30, 2010

Report

Report Number
1824206-2010-09649
Event Type
Malfunction
Date Received
August 30, 2010
Date of Event
August 5, 2010
Report Date
August 5, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH WAS UNABLE TO ADJUST THE BRAKE. HE FOUND THE BRAKE CABLE BINDING BETWEEN PLATE AND OLD STYLE BEARINGS. THE TECH REPLACED THE BEARINGS TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE CASTERS ARE NOT HOLDING IN BRAKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY BED AC POWERED HOSPITAL BED FNL HILL-ROM INC. 840

Patients

Seq Age Sex Outcome Treatment
1