FDA Adverse Event
Malfunction
Summary report: N
CENTURY BED
MDR report key: 1823341
·
Received August 30, 2010
Report
- Report Number
- 1824206-2010-09649
- Event Type
- Malfunction
- Date Received
- August 30, 2010
- Date of Event
- August 5, 2010
- Report Date
- August 5, 2010
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH WAS UNABLE TO ADJUST THE BRAKE. HE FOUND THE BRAKE CABLE BINDING BETWEEN PLATE AND OLD STYLE BEARINGS. THE TECH REPLACED THE BEARINGS TO REPAIR THE BED.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE CASTERS ARE NOT HOLDING IN BRAKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTURY BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM INC. | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |