FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 1823165 · Received September 1, 2010

Report

Report Number
1226181-2010-00117
Event Type
Malfunction
Date Received
September 1, 2010
Date of Event
August 18, 2010
Report Date
August 19, 2010
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
MMI
PMA / PMN Number
K944093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS THE CHEM WASH LINE WAS NOT CONNECTED TO THE WASH PROBE. THE CUSTOMER CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

FOUR FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON PATIENT SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED AND NEGATIVE RESULTS WERE OBTAINED. ONE PATIENT RECEIVED A CARDIAC CATHETERIZATION. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.