FDA Adverse Event
Malfunction
Summary report: N
DIMENSION® CLINICAL CHEMISTRY SYSTEM
MDR report key: 1823165
·
Received September 1, 2010
Report
- Report Number
- 1226181-2010-00117
- Event Type
- Malfunction
- Date Received
- September 1, 2010
- Date of Event
- August 18, 2010
- Report Date
- August 19, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS THE CHEM WASH LINE WAS NOT CONNECTED TO THE WASH PROBE. THE CUSTOMER CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
FOUR FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON PATIENT SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED AND NEGATIVE RESULTS WERE OBTAINED. ONE PATIENT RECEIVED A CARDIAC CATHETERIZATION. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC | RXLMAXHM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |