INTEGRATED APD SET W/CASSETTE3-PRONG
Report
- Report Number
- 1423500-2010-03066
- Event Type
- Malfunction
- Date Received
- September 1, 2010
- Date of Event
- August 8, 2010
- Report Date
- August 8, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE AVAILABILITY IS UNKNOWN AT THIS TIME. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA NUMBER (B)(4).
(B)(4). THIS COMPLAINT FOR A LOW DRAIN VOLUME WITH A REPORT OF AIR IN THE LINE WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE CAUSE WAS UNDETERMINED. THE BATCH REVIEW WAS NOT PERFORMED BECAUSE THE LOT NUMBER IS UNKNOWN. NO EVALUATION WILL BE PERFORMED. PRODUCT SURVEILLANCE FOLLOWED UP WITH THE PATIENT'S NURSE REGARDING THE ALARM. THE NURSE STATED THE PATIENT HAD NOT REPORTED THE INCIDENT TO HER, BUT THAT HE HAS HAD NO INJURIES OR ILLNESSES. THE PATIENT CONTINUES TO USE THE HOMECHOICE FOR PD THERAPY TO DATE.
A CUSTOMER CONTACTED BAXTER'S (B)(4) REGARDING A LOW DRAIN VOLUME ALARM ON THE HOMECHOICE (HC) DURING INITIAL DRAIN. DURING TROUBLESHOOTING THE HOME PATIENT (HP) REPORTED AIR IN THE PATIENT LINE AND REQUESTED TO BYPASS. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THAT IF THERE WAS AIR IN THE LINE HE COULD NOT BYPASS TO FILL. TSR SUGGESTED TO BEGIN AGAIN WITH NEW SUPPLIES AND TO CALL BACK WITH ANY QUESTIONS. HP WILL START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |