FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 18228905 · Received November 29, 2023

Report

Report Number
1221359-2023-01684
Event Type
Malfunction
Date Received
November 29, 2023
Date of Event
January 1, 2023
Report Date
December 8, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D4- UDI: (B)(4). DATE OF EVENT PROVIDED IN SECTION B3 IS AN APPROXIMATION, WAS NOT PROVIDED BY CONSUMER. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : DEVICE DISCARDED; SINGLE-USE DEVICE.

Additional Manufacturer Narrative · 0

D4- UDI: (B)(4). DATE OF EVENT PROVIDED IN SECTION B3 IS AN APPROXIMATION, WAS NOT PROVIDED BY CONSUMER. TESTING WAS PERFORMED IN TRIPLICATE AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 203262 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 195-160 / LOT 203262 AND DEVICE PART NUMBER 195-430H / LOT 198241. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. THE REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED, UNCONFIRMED AND CONFLICTING RESULTS) RELATED TO KIT LOT 203262 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : DEVICE DISCARDED; SINGLE-USE DEVICE.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON AN UNKNOWN DATE. THIS MFR. REPORT ADDRESSES RESULT ONE (1) OF TWO (2). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST ASSAY PERFORMED ON AN UNKNOWN DATE. ADDITIONAL TESTING VIA IHEALTH WAS PERFORMED TWICE ON AN UNKNOWN DATE AND GENERATED A POSITIVE RESULT BOTH TIMES. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON AN UNKNOWN DATE. THIS MFR. REPORT ADDRESSES RESULT ONE (1) OF TWO (2). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST ASSAY PERFORMED ON AN UNKNOWN DATE. ADDITIONAL TESTING VIA IHEALTH WAS PERFORMED TWICE ON AN UNKNOWN DATE AND GENERATED A POSITIVE RESULT BOTH TIMES. NO ADDITIONAL INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
982653 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 203262 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown