FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18228760 · Received November 29, 2023

Report

Report Number
2647876-2023-00600
Event Type
Malfunction
Date Received
November 29, 2023
Date of Event
September 18, 2023
Report Date
November 28, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: CATALOG: 442021, BATCH NO.: 3130610 CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION/RETURNED SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. NO TREND IDENTIFIED FOR BATCH. COMPLAINT IS UNCONFIRMED BASED ON RETENTION/RETURNED SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) THERE WAS ONE MOLECULAR FALSE POSITIVE OF A TESTING BOTTLE. NO PATIENT IMPACT REPORTED. THIS IS REPORT 2 OF 2. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTING C TROPICALIS MOLECULAR FALSE POSITIVE FOR PRODUCT 442021 LOT NO. 3130610 WAS THIS ISSUE INVOLVING PATIENT OR NON-PATIENT SAMPLES (QC, VALIDATION TESTING OR PROFICIENCY SAMPLES)? ONE PATIENT WAS THIS ISSUE INVOLVING PATIENT OR NON-PATIENT SAMPLES (QC, VALIDATION TESTING OR PROFICIENCY SAMPLES)? ONE PATIENT AND ONE BLANK BOTTLE FOR TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67890 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3130610 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown