FDA Adverse Event Malfunction Summary report: N

TLOC 133 OFFSET INSERTER ASSY

MDR report key: 18226808 · Received November 29, 2023

Report

Report Number
0001825034-2023-02787
Event Type
Malfunction
Date Received
November 29, 2023
Date of Event
November 8, 2023
Report Date
March 19, 2024
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
UDI-DI
00880304509719
PMA / PMN Number
K120030
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. ONE TLOC 133 OFFSET INSERTER ASSY ITEM# 51-222224 LOT# 057777 WAS RETURNED AND EVALUATED. UPON VISUAL INSPECTION THE TIP OF THE DEVICE HAD FRACTURED FROM THE DEVICE. THERE ARE INDENTATIONS TO THE STRIKE PLATE. A REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. THE EVENT IS CONFIRMED VIA RETURNED PRODUCT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

DURING IMPACTION OF THE FEMORAL COMPONENT, THE INSERTER BROKE. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Description of Event or Problem · 0

NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
982526 TLOC 133 OFFSET INSERTER ASSY INSTRUMENT, HIP LPH ZIMMER BIOMET, INC. N/A 057777 00880304509719

Patients

Seq Age Sex Outcome Treatment
1 NA Male