FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 36 CM

MDR report key: 1822667 · Received November 27, 2007

Report

Report Number
1527736-2007-07919
Event Type
Malfunction
Date Received
November 27, 2007
Report Date
September 5, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR". EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. THE MFG RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING LAP SIGMOID PROCEDURE AFTER A FEW MINUTES NO FUNCTION. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 36 CM LFL ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA C9D017

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR