FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1822636 · Received November 27, 2007

Report

Report Number
1527736-2007-07933
Event Type
Malfunction
Date Received
November 27, 2007
Date of Event
October 30, 2007
Report Date
October 30, 2007
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED; HOWEVER, A SQUEALING NOISE WAS HEARD. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AN NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO AN UNK PROCEDURE THE HANDPIECE WAS FOUND TO HAVE A LOOSE MOUNT. NO PT CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO SURGERY, INC (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1