FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1822636
·
Received November 27, 2007
Report
- Report Number
- 1527736-2007-07933
- Event Type
- Malfunction
- Date Received
- November 27, 2007
- Date of Event
- October 30, 2007
- Report Date
- October 30, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED; HOWEVER, A SQUEALING NOISE WAS HEARD. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AN NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO AN UNK PROCEDURE THE HANDPIECE WAS FOUND TO HAVE A LOOSE MOUNT. NO PT CONSEQUENCES WERE REPORTED.