FDA Adverse Event Malfunction Summary report: N

HARMONIC ACE 23 CM

MDR report key: 1822587 · Received November 29, 2007

Report

Report Number
1527736-2007-08044
Event Type
Malfunction
Date Received
November 29, 2007
Date of Event
September 14, 2007
Report Date
October 15, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT WHEN ATTEMPTING TO ACTIVATE DEVICE A ON A GEN04 GAVE AN ERROR CODE 5. VISUAL EXAMINATION FOUND AN AREA OF THE BLADE THAT WAS DISCOLORED DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. FURTHER ANALYSIS FOUND A CRACK PROPAGATING FROM THE HEAT-AFFECTED AREA OF THE BLADE. THE ANALYSIS RESULTS FOUND THAT WHEN ATTEMPTING TO ACTIVATE DEVICE B ON A GEN04 GAVE AN ERROR CODE 5. VISUAL EXAMINATION FOUND AN AREA OF THE BLADE THAT WAS DISCOLORED DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. FURTHER ANALYSIS FOUND A CRACK PROPAGATING FROM THE HEAT-AFFECTED AREA OF THE BLADE. THIS FAILURE IS CAUSED DUE TO ACTIVATION OF THE DEVICE WHILE CLAMPED WITHOUT TISSUE BEING PRESENT. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT." WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAR PROCEDURE AN ERROR OCCURRED VERY OFTEN (CODE 5 WITH A GEN04). ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC ACE 23 CM LFL ETHICON ENDO SURGERY, INC. (CINCINNATI) NA D4HD4A

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR