FDA Adverse Event Injury Summary report: N

TALENT THORACIC STENT GRAFT SYSTEM

MDR report key: 1822562 · Received August 27, 2010

Report

Report Number
2953200-2010-01650
Event Type
Injury
Date Received
August 27, 2010
Date of Event
July 28, 2010
Report Date
July 28, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: RESULTS: ARTERIAL TRAUMA/DISSECTION/PERFORATION, ARTERIAL AND VENOUS OCCLUSION; VESSEL DIAMETER. CONCLUSIONS: VESSEL DIAMETER.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 6.3 CM THORACIC AORTIC ANEURYSM. THE RIGHT ILIAC ARTERY WAS 8.5 MM IN DIAMETER AND THE LEFT ILIAC ARTERY WAS 9 MM IN DIAMETER. THE VESSELS WERE MILDLY TORTUOUS AND MILDLY CALCIFIED. IT WAS REPORTED THAT WHEN THE DISTAL MAIN WAS INSERTED INTO THE PT, IT WOULD NOT TRACK UP THE PT'S ANATOMY AND RESULTED IN A SMALL ILIAC TEAR. THE ILIAC ARTERY WAS REPAIRED WITH A PATCH. THE DELIVERY CATHETER WAS REMOVED FROM THE PT AND ANOTHER DISTAL MAIN WAS USED TO COMPLETE THE CASE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT THORACIC STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00453166

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention