FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT

MDR report key: 1822560 · Received August 27, 2010

Report

Report Number
2953200-2010-01653
Event Type
Injury
Date Received
August 27, 2010
Date of Event
July 30, 2010
Report Date
July 30, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: RESULTS: ENDOLEAK, MIGRATION, DISEASE PROGRESSION, AND SHORT AORTIC NECK. CONCLUSION: DISEASE PROGRESSION, AND SHORT AORTIC NECK.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR ENDOVASCULAR TREATMENT OF A 6 CM ABDOMINAL AORTIC ANEURYSM APPROX 24 MONTHS AGO. VESSEL MORPHOLOGY AT THE TIME OF IMPLANT WAS REPORTED AS THE AORTIC NECK CHANGES FROM 21 MM IN DIAMETER AT THE RENAL ARTERIES AND 30 MM DISTALLY THE DIAMETER IS 29 MM IN DIAMETER. CURRENTLY, THE AORTIC NECK IS SHORT MEASURING 1 - 1.5 CM AND IT WAS FUNNEL SHAPED. IT WAS REPORTED THAT A RECENT CT DEMONSTRATED THAT THE STENT GRAFT HAS MIGRATED AND IS 1 CM BELOW THE RENAL ARTERIES WITH A TYPE 1 ENDOLEAK PRESENT. THE PT WILL BE TREATED IN THE NEXT MONTH. NO ADDITIONAL CLINICAL SEQUELAE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT MIH MEDTRONIC CARDIOVASCULAR NA V00113705

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention