PENTAX
Report
- Report Number
- 9610877-2023-61386
- Event Type
- Malfunction
- Date Received
- November 29, 2023
- Date of Event
- November 16, 2023
- Report Date
- November 29, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB15-J10 IS AVAILABLE IN THE USA WITH A 510K NUMBER K200678. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE DISTAL END WITH CCD MODULE FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE DISTAL END WITH CCD MODULE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE U/D KNOB BROKEN, THE BENDING RUBBER PERFORATED, THE LCB DISTAL COVER GLASS CRACKED, AND THE INSERTION FLEXIBLE TUBE TWISTED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 79221 | PENTAX | VIDEO BRONCHOSCOPE | EOQ | HOYA CORPORATION PENTAX TOKYO OFFICE | EB11-J10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |