FDA Adverse Event Summary report: N

STAT PADZ

MDR report key: 1822470 · Received August 25, 2010

Report

Report Number
1822470
Date Received
August 25, 2010
Date of Event
July 29, 2010
Report Date
August 25, 2010
Manufacturer
ZOLL MEDICAL CORP
Product Code
MKJ
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

WHEN THE PLUG WAS PULLED OUT OF THE OUTLET, THE METALLIZED ZOLL STAT PADZ BAG TOUCHED THE PLUG PRONGS AND SHORTED OUT THE OUTLET AND DAMAGED THE ALARIS PLUG. THERE WAS NO DAMAGE TO THE ALARIS UNIT ITSELF. ------------------------------------------------------------ZOLL STAT PAD WAS HANGING AT HOB HEAD OF BED FROM METAL CLIP ON THE WALL AND WAS RIGHT NEXT TO A RED OUTLET THAT HAD AN IV PLUGGED INTO THE RED OUTLET. AS THE PLUG WAS PULLED OUT OF WALL, IT MUST HAVE TOUCHED THE PLUG TO THE ZOLL PAD PACKAGING AND SPARKS, POPS AND SMOKE CAME OUT OF THE OUTLET. ELECTRICAL OUTLET EXAMINED BY BIOMED AND DETERMINED TO HAVE NO EFFECT TO THE OUTLET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STAT PADZ AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) MKJ ZOLL MEDICAL CORP * *

Patients

Seq Age Sex Outcome Treatment
1 *