AVALON ELITE
Report
- Report Number
- 2032228-2010-00005
- Event Type
- Injury
- Date Received
- August 26, 2010
- Date of Event
- June 24, 2010
- Report Date
- August 24, 2010
- Manufacturer
- AVALON LABORATORIES LLC
- Product Code
- DWF
- PMA / PMN Number
- 081820
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEVICE HISTORY RECORD AND INSPECTION DOCUMENT (LOTS 2131608, 2131609, 2131610, AND 2131955) RELATED TO THE PRODUCT USED WAS REVIEWED AND FOUND NO NON CONFORMITIES. THE 13FR BI-CAVAL LUMEN CATHETER ARE INSPECTED 100% IN PRODUCTION. THE IFU WARNS THAT INCORRECT INSERTION CAN CAUSE DAMAGE TO THE VESSELS AND/OR HEART STRUCTURES.
BRIEF DESCRIPTION: TETRALOGY OF FALLOT WITH PERSISTENT PULMONARY HYPERTENSION CANNULATED FOR VV ECMO BY (B)(6). CANNULA INSERTED WITH FLUOROSCOPIC GUIDANCE. NO IMMEDIATE COMPLICATION APPARENT. CANNULA APPEARED IN GOOD POSITION ON INITIAL CHEST X-RAY BUT SUBSEQUENTLY MOVED. ECHOCARDIOGRAPHY SHOWED CANNULA TIP TO BE WITHIN THE CORONARY SINUS AND WAS REPOSITIONED AND A TRIAL OFF COMMENCED AFTER 40 HOURS OF ECMO SUPPORT. AS IS OUR PRACTICE ON COMMENCING TRIAL OFF THE FLOW WAS INCREASED. THERE WAS THEN A CARDIAC ARREST AND REPEAT ECHO REVEALED A PERICARDIAL EFFUSION. REMEDIAL ACTION TAKEN AND BY WHOM UPON THE OCCURRENCE OF THE EVENT. THE CHEST WAS OPENED THROUGH A MEDIAN STERNOTOMY AND A HOLE IN THE RIGHT ATRIUM WAS IDENTIFIED JUST ANTERIOR TO THE ENTRANCE OF THE INFERIOR VENA CAVA. THIS WAS REPAIRED WITH A SINGLE PROLENE SUTURE AND AS THE PT WAS STABLE OFF ECMO THE CANNULA WAS REMOVED AND THE CHEST CLOSED. PT OUTCOME: THE PT REMAINED STABLE AND WAS DISCHARGED FROM THE HOSPITAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AVALON ELITE | 13 FR BI-CAVAL LUMEN CATHETER | DWF | AVALON LABORATORIES LLC | 13 FR AVALON ELITE | 090216017953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2 DA | Life Threatening |