FDA Adverse Event Malfunction Summary report: N

LONG 45 ENDOCUTTER

MDR report key: 1822382 · Received August 16, 2007

Report

Report Number
1527736-2007-05441
Event Type
Malfunction
Date Received
August 16, 2007
Date of Event
July 13, 2007
Report Date
July 16, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH THE CLAMPING MECHANISM DAMAGED AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED FULLY LOADED WITH STAPLES. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND YOKE TEETH WERE FOUND BROKEN. THE RETURNED CARTRIDGE WAS LOADED INTO A TEST DEVICE AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE TEETH BECAME BROKEN, THIS DAMAGE CAN OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE CLOSING TRIGGER WHEN THE JAWS ARE CLAMPED ON TISSUE THICKER THAN INDICATED. IN ADDITION, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE, THE DEVICE DID NOT FIRE. A NEW DEVICE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LONG 45 ENDOCUTTER GDW ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA D4H256

Patients

Seq Age Sex Outcome Treatment
1