HARMONIC ACE 36 CM
Report
- Report Number
- 1527736-2007-05530
- Event Type
- Malfunction
- Date Received
- August 17, 2007
- Date of Event
- September 12, 2006
- Report Date
- September 14, 2006
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
(B)(4). BURNT AND CRACKED. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT WHEN ATTEMPTING TO ACTIVATE THE DEVICE ON A GEN04 GAVE AN ERROR CODE 5. VISUAL EXAMINATION FOUND AN AREA OF BLADE THAT WAS DISCOLORED DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. FURTHER ANALYSIS FOUND A CRACK PROPAGATING FROM THE HEAT-AFFECTED AREA OF THE BLADE. THIS FAILURE IS CAUSED DUE TO ACTIVATION OF THE DEVICE WHILE CLAMPED WITHOUT TISSUE BEING PRESENT. FOR THIS REASON, THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT." WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AN NO ANOMALIES WERE FOUND DURING THE MFG PROCESS.
IT WAS REPORTED THAT PRIOR TO A LAVH PROCEDURE, THE NURSE WAS DOING NORMAL CALIBRATION OF INSTRUMENT, IN GENERATOR TEST IN PROGRESS NORMALLY, BUT AFTER THAT ONLY ERROR SOUND. NO ERROR CODE ON DISPLAY. THEY TRY TO CALIBRATE INSTRUMENT AGAIN BUT SAME PROBLEM. ALSO THE INSTRUMENT WAS MAKING VERY HIGH SOUND. HAND PIECE AND GENERATOR TESTED ACCORDING TO ADVANCED TROUBLE SHOOTING MANUAL. BOTH OK. NOT NOTED HOW CASE WAS COMPLETED. NO PT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO SURGERY, INC (CINCINNATI) | NA | C4ER7U |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |