FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE

MDR report key: 1822365 · Received August 16, 2007

Report

Report Number
1527736-2007-05438
Event Type
Malfunction
Date Received
August 16, 2007
Date of Event
July 3, 2007
Report Date
July 3, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DAMAGED ARTICULATION GEAR TEETH. EVAL SUMMARY: THE ANALYSIS SHOWED THAT THE DEVICE WAS RECEIVED WITH THE ARTICULATION MECHANISM DAMAGED. THE DEVICE WAS RECEIVED WITH A CARTRIDGE LOADED IN THE DEVICE; THE CARTRIDGE WAS RECEIVED FULLY LOADED WITH STAPLES. THE RETURNED CARTRIDGE WAS TESTED FOR FUNCTIONALITY WITH A TEST DEVICE AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST CARTRIDGE ON THE 3 ARTICULATED POSITIONS; WHEN FIRED TO THE LEFT AND CENTER THE STAPLES WERE MALFORMED DUE TO THE DAMAGE ARTICULATION MECHANISM, HOWEVER, WHEN FIRE IN THE RIGHT POSITION THE STAPLE FORMATION WAS CONFORMING. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE ARTICULATION GEAR TEETH WERE FOUND BROKEN. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A TAH/BSO PROCEDURE, THE ARTICULATION BAND BROKE. THE STAPLER WOULD NOT ARTICULATE ANYMORE. A NEW DEVICE WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS COMPACT FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FOR OPEN PROCE GDW ETHICON ENDO SURGERY, INC. (CINCINNATI) NA D4GV1N

Patients

Seq Age Sex Outcome Treatment
1