FDA Adverse Event
Malfunction
Summary report: N
HS HAND CONTROLLED CURVED SHEA
MDR report key: 1822351
·
Received December 6, 2007
Report
- Report Number
- 1527736-2007-08356
- Event Type
- Malfunction
- Date Received
- December 6, 2007
- Date of Event
- June 14, 2007
- Report Date
- July 19, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- GEI
- PMA / PMN Number
- K002981
- Removal / Correction Number
- NA
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): INFO NOT PROVIDED DURING INITIAL CONTACT. EVAL SUMMARY: THE DEVICE WAS RECEIVED IN GOOD CONDITION. NO VISIBLE DAMAGE WAS NOTED. THE HAND-ACTIVATION CONTACTS WERE IN GOOD CONDITION. THE DEVICE WAS TESTED ON A GENERATOR AND IT WAS FOUND THAT THE HAND CONTROL SWITCH ASSEMBLY BUTTONS WERE NOT FUNCTIONAL. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS HAND CONTROLLED CURVED SHEA | GEI | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |