HYDRO X-FLOW CATHETER SILICONE WITH HYDROGEL COATING
Report
- Report Number
- 9610711-2023-00255
- Event Type
- Malfunction
- Date Received
- November 28, 2023
- Report Date
- June 13, 2024
- Manufacturer
- COLOPLAST A/S
- Product Code
- EZL
- UDI-DI
- 03600040260753
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ACCORDING TO THE COMPLAINT DESCRIPTION, THE LOT NUMBER IS AVAILABLE BUT NOT THE SAMPLE. AFTER RECEIVING THIS COMPLAINT, WE SEARCHED FOR OTHER COMPLAINTS AND FOUND NONE ON THE LOT NUMBER 9240226. CHECKING THE QUALITY DATABASES REVEALED ONE NON-CONFORMITY (NC) POTENTIALLY IN RELATION TO THE DESCRIBED DEFECT AND A CORRECTIVE AND PREVENTIVE ACTION (CAPA): - NC (B)(4): "PB BALLONS (BULLES + AGGLUTINÉS)" OPENED IN 16 JANUARY 2023. FOR THIS NC, THE USED OF CLUMPED BALLOONS SHOULD BE RESPONSIBLE OF SOME WEAKNESS ON THE BALLOON AND BEING RESPONSIBLE OF BURSTING. - MONITORING CAPA-000152: "BALLOON ISSUES ON FOLYSIL AND SILICONE PROSTATIC CATHETERS". THE TRENDING FOR BALLOON BURSTING IS SPECIFICALLY MONITORED. A SIMILAR CASE STUDY WAS PERFORMED BASED ON ITEM NUMBER XB63, DEFECT BALLOON DEFLATION. BALLOON BURST; DEFECTIVE BALLOON OVER LAST FOUR YEAR, 42 SIMILAR CASES WERE FOUND.
ACCORDING TO THE AVAILABLE INFORMATION UPON INSERTION OF THE CATHETER, THE BALLOON BURST ON INFLATION AT 30ML. NO CLINICAL CONSEQUENCE FOR THE PATIENT.
ACCORDING TO THE AVAILABLE INFORMATION UPON INSERTION OF THE CATHETER, THE BALLOON BURST ON INFLATION AT 30ML. NO CLINICAL CONSEQUENCE FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 532167 | HYDRO X-FLOW CATHETER SILICONE WITH HYDROGEL COATING | INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL | EZL | COLOPLAST A/S | 9240226_XB63181002 | 03600040260753 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |