FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1822115 · Received August 24, 2010

Report

Report Number
2023826-2010-00824
Event Type
Malfunction
Date Received
August 24, 2010
Date of Event
July 29, 2010
Report Date
July 29, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE). RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED ONLY A HAPTIC AND A SMALL PIECE OF THE OPTIC WERE RECEIVED. THERE WAS EVIDENCE OF A CLEAR SURGICAL RESIDUE. CONCLUSION: A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVALUATION OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND A HAPTIC BROKE UPON INSERTION. THE LENS WAS REMOVED WITHOUT ANY PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 UNK CARTRIDGE: MODEL CQ CARTRIDGE-FP, LOT NUMBER UNK| INJECTOR: MODEL MSI-PM, LOT NUMBER UNK