FDA Adverse Event Injury Summary report: N

TSX IMPLANT, 4.1MMD, 13MML

MDR report key: 18220692 · Received November 28, 2023

Report

Report Number
0001038806-2023-02276
Event Type
Injury
Date Received
November 28, 2023
Date of Event
October 23, 2023
Report Date
June 4, 2024
Manufacturer
PBG 3I
Product Code
DZE
UDI-DI
00844868045113
PMA / PMN Number
K220978
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT (B)(4). A1: PATIENT IDENTIFIER IS NOT PROVIDED / UNKNOWN. A2: PATIENT AGE IS NOT PROVIDED / UNKNOWN. A4: PATIENT WEIGHT IS NOT PROVIDED / UNKNOWN. E1: INITIAL REPORTER¿S TITLE, FIRST NAME, LAST NAME AND EMAIL ADDRESS ARE NOT PROVIDED / UNKNOWN.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO REPORT ADDITIONAL INFORMATION. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.)/PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO MEDICAL CONDITIONS/PATIENT HABITS AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION AND NO ASSOCIATION BETWEEN THE DENTAL IMPLANT AND BONE LOSS WAS FOUND THAT CAN EXPLAIN THESE EVENTS. THEREFORE, THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS PROBLEMS REPORTED IN ASSOCIATION WITH IMPLANT FAILURE HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #34 WAS REMOVED DUE TO BONE LOSS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION.

Description of Event or Problem · 0

NO ADDITIONAL OR CORRECTED INFORMATION TO REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
602311 TSX IMPLANT, 4.1MMD, 13MML DENTAL IMPLANT DZE PBG 3I 1262131 00844868045113

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose Required Intervention