ICL (IMPLANTABLE COLLAMER LENS)
Report
- Report Number
- 2023826-2010-00817
- Event Type
- Injury
- Date Received
- August 20, 2010
- Date of Event
- August 5, 2010
- Report Date
- August 10, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4)
EVALUATION METHOD: LENS WORK ORDER SEARCH, MEDICAL REVIEW. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND ONE HAPTIC BROKEN. THE LENS WAS RETURNED DRY AND THERE WAS BLUE INK ON THE LENS. PER MEDICAL REVIEW - ACCORDING TO USE FMEA (FAILURE MODES AND EFFECT ANALYSIS) IT HAS BEEN DETERMINED THAT EXCESSIVE VAULTING IS A CONSEQUENCE OF A WRONG LENS USE FAILURE MODE (I.E. IMPROPER WHITE TO WHITE MEASUREMENT, VARIABILITY OF THE WHITE TO WHITE MEASUREMENTS BASED UPON DIFFERENT TECHNIQUES UTILIZED, IMPROPER SULCUS TO SULCUS MEASUREMENT (IF UBM USED), AND PATIENT CONDITION; POOR CORRELATION OF WHITE TO WHITE MEASUREMENT AND LENGTH OF CILIARY SULCUS IN AN INDIVIDUAL CASE; IRREGULAR CILIARY SULCUS OR CILIARY SULCUS CYST). SEVERAL CONDITIONS MAY ARISE FROM THIS EVENT (I.E. PUPILLARY BLOCK, MALIGNANT GLAUCOMA, INCREASED IOP ETC.). TO PREVENT ANY OF THESE COMPLICATIONS FROM OCCURRING, THE MANUFACTURER RECOMMENDS THAT THE LENS BE EXPLANTED AND/OR EXCHANGED WITH THE RIGHT SIZE ONCE THE SURGEON DETERMINES THAT THIS CONDITION MAY AFFECT THE OUTCOME OF THE PATIENT'S VISION. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, MEDICAL REVIEW AND EVALUATION OF THE RETURNED PRODUCT, THE MOST PROBABLE ROOT CAUSE OF THE EVENT IS EXCESSIVE VAULT SECONDARY TO A LONG LENS LENGTH. (B)(4).
THE REPORTER STATED THE SURGEON IMPLANTED A 12.5MM ICM125V4 IMPLANTABLE COLLAMER LENS IN THE PT'S RIGHT EYE (OD) ON (B)(6)2010. THE LENS WAS EXPLANTED ON (B)(6)2010 DUE TO EXCESSIVE VAULTING. THE ICL WAS EXCHANGED FOR A SHORTER LENS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | ICM125V4 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | INJECTOR: MODEL AND LOT NUMBER UNK| CARTRIDGE: MODEL SFC-45 FP - LOT NUMBER 1255130| LOT NUMBER 1253987| FOAM TIP PLUNGER: MODEL FOAM TIP PLUNGER |