FDA Adverse Event Malfunction Summary report: N

CYAN ADP HIGH SPEED ANALYZER

MDR report key: 1821919 · Received August 31, 2010

Report

Report Number
1061932-2010-00131
Event Type
Malfunction
Date Received
August 31, 2010
Date of Event
August 2, 2010
Report Date
August 31, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
GKZ
PMA / PMN Number
RUO
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) CHECKED THE OUTLET POWER AND IT MEASURED 117VAC. PER PRODUCT LABELING, THE POWER REQUIREMENTS ARE 100V - 230V, 600VA (W), 50/60 HZ. A NEW WORKSTATION WAS ORDERED.

Description of Event or Problem · 1

DURING INSTALLATION OF THE UNIT, THE FIELD SERVICE ENGINEER (FSE) REPORTED SMOKE COMING FROM THE POWER SUPPLY WHEN IT WAS PLUGGED INTO THE WALL OUTLET. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CYAN ADP HIGH SPEED ANALYZER FLOW CYTOMETER GKZ BECKMAN COULTER INC. NA

Patients

Seq Age Sex Outcome Treatment
1