FDA Adverse Event
Malfunction
Summary report: N
CYAN ADP HIGH SPEED ANALYZER
MDR report key: 1821919
·
Received August 31, 2010
Report
- Report Number
- 1061932-2010-00131
- Event Type
- Malfunction
- Date Received
- August 31, 2010
- Date of Event
- August 2, 2010
- Report Date
- August 31, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- GKZ
- PMA / PMN Number
- RUO
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE ENGINEER (FSE) CHECKED THE OUTLET POWER AND IT MEASURED 117VAC. PER PRODUCT LABELING, THE POWER REQUIREMENTS ARE 100V - 230V, 600VA (W), 50/60 HZ. A NEW WORKSTATION WAS ORDERED.
Description of Event or Problem · 1
DURING INSTALLATION OF THE UNIT, THE FIELD SERVICE ENGINEER (FSE) REPORTED SMOKE COMING FROM THE POWER SUPPLY WHEN IT WAS PLUGGED INTO THE WALL OUTLET. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CYAN ADP HIGH SPEED ANALYZER | FLOW CYTOMETER | GKZ | BECKMAN COULTER INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |