FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE
Report
- Report Number
- 6000001-2010-02905
- Event Type
- Malfunction
- Date Received
- August 31, 2010
- Date of Event
- July 8, 2010
- Report Date
- July 20, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH FAILURE CODE 2 WAS CONFIRMED DURING PRODUCT EVALUATION IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY THE CHANNEL 1 DOOR ASSEMBLY BEING BROKEN IN THE LATCH AREA. THE CHANNEL 1 DOOR ASSEMBLY WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH FAILURE CODE 2, WHICH IS A DOWNSTREAM OCCLUSION FAILURE CODE. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS FACILITY WAS NOT WILLING TO BE CONTACTED FOR ANY FURTHER INFORMATION. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. DURING PRODUCT EVALUATION, FAILURE CODE 2 WAS CONFIRMED TO BE CAUSED BY THE CHANNEL 1 DOOR ASSEMBLY BEING BROKEN IN THE LATCH AREA, INDICATING FAILURE CODE 2 WAS A FALSE OCCLUSION ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |