FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1821756 · Received August 26, 2010

Report

Report Number
9615739-2010-00207
Event Type
Malfunction
Date Received
August 26, 2010
Date of Event
August 2, 2010
Report Date
August 2, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THAT THE RIGHT UCM BOARD HEAD DOWN BUTTON WAS STUCK. HE REPLACED THE RIGHT UCM BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE HEAD OF BED WOULD RISE AND THEN LOWER ON ITS OWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC POWERED HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1