SEALPTFE
Report
- Report Number
- 9612515-2010-00002
- Event Type
- Other
- Date Received
- August 27, 2010
- Date of Event
- July 30, 2010
- Report Date
- August 27, 2010
- Manufacturer
- VASCUTEK LTD.
- Product Code
- DSY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
METHOD - DEVICE TO BE RETURNED TO VASCUTEK FOR EVAL. RESULTS: REVIEW OF MFG & QC RECORDS. THE MFG AND QC RECORDS FOR THE BATCH HAVE BEEN REVIEWED AND THERE IS NOTHING TO SUGGEST A PROBLEM WITH THE BATCH. THE GRAFT IN QUESTION WAS CONTAINED IN A BATCH OF 39 UNITS WHICH WERE DISPATCHED IN (B)(6) 2009. WE HAVE NOT RECEIVED ANY OTHER SIMILAR REPORTS FROM THIS BATCH. THE USAGE RATE OF THIS PARTICULAR MODEL IS SUCH THAT IT MAY BE ASSUMED THE MAJORITY HAVE NOW BEEN IMPLANTED. THIS TYPE OF SWELLING, DUE TO SEROMA FORMATION, IS A KNOWN COMPLICATION WITH EPTFE VASCULAR PROSTHESIS. THE INCIDENCE OF SEROMA IS HIGHER IN EXTRA-ANATOMIC GRAFTS THAN CONVENTIONAL ABDOMINAL IMPLANTS. BLUMENBERG REPORTS OF A SURVEY OF 279 CASES WHERE PEIRGRAFT SEROMAS WERE IDENTIFIED. FIFTY FOUR PERCENT OF CASES INVOLVED KNITTED DACRON AND 34% INVOLVED PTFE. HE STATES THAT GRAFT REPLACEMENT PROVIDED A 92% CURE RATE. THE ASSOCIATION OF SEROMA WITH IMPLANT LOCATION RATHER THAN THE GRAFT TYPE WAS ALSO SHOWN BY BLUMENBERGL. FOUR PERCENT OF THE SEOMAS HE DESCRIBED WERE WITH AXILLOAXILLARY GRAFTS WITH EPTFE GRAFTS, ULTRA-FILTRATION OF SERUM THROUGH THE GRAFT WALL SEEMS THE LIKELY CAUSE OF SEROMA. IT IS POSSIBLE THAT SOME FLUID COLLECTS AROUND MANY GRAFTS, BUT IS MORE APPARENT IN TUNNELED DEVICES THAN THOSE IN THE ABDOMEN. IT IS ALSO POSSIBLE THAT SOME ABNORMALITY IN THE HEALING PROCESS ALLOWS SERUM TO PERMEATE THROUGH THE GRAFT WALL. PRICOLO TO THE SECOND POWER FOUND THAT SEROMA FLUID INHIBITED FIBROLBLAST GROWTH. IN SOME CASES, SEROMA HAS ONLY RESOLVED WHEN THE GRAFT TYPE WAS CHANGED, I.E., EPTFE TO POLYESTER, OR VICE-VERSA, SUGGESTING AN ALLERGENIC MECHANISM. SYMPTOMS WENT AWAY WHEN GRAFT WAS EXPLANTED. VASCUTEK WILL WRITE TO THE DOCTOR CONCERNED PROVIDING THE INFO GIVEN IN THE REFERENCES AND RESULTS OF THE GRAFT INVESTIGATION. VASCUTEK WILL FOLLOW-UP WITH A FINAL REPORT. BLUMENBERG RM, GELFAND ML AND DALE WA. PERIGRAFT SEROMAS COMPLICATING ARTERIAL GRAFTS. SURGERY 1985;97(2):194-203. PRICOLO VE, POTENTI F, SODERBERG CH. EFFECT OF PERIGRAFT SEROMA FLUID ON FIBROBLAST PROLIFERATION IN VITRO. ANNALS OF VASCULAR SURGERY 1991;5(5):462-466. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE EVENT OCCURRED AT (B)(6) IN (B)(6) USA. THE PT HAS END STAGE RENAL DISEASE AND HAD PLACEMENT OF VASCUTEK SEALPTFE GELATIN SEALED VASCULAR GRAFT FOR HEMODIALYSIS. POSTOPERATIVELY THE PT EXPERIENCED UNUSUAL SWELLING OVER THE INCISION OVER THE ARTERIAL ANASTOMOSIS. THE CAUSE WAS A COLLECTION OF SERUM. THE WOUND WAS TREATED WITH REPEATED ASPIRATION BUT THE FLUID KEPT RE-ACCUMULATING. THEN WOUND WAS RE-EXPLORED AND HE WAS FOUND TO HAVE EXCESSIVE ULTRAFILTRATION (WEEPING) THROUGH THE GRAFT ADJACENT TO THE ARTERIAL ANASTOMOSIS AND A 50 ML COLLECTION OF SERUM. THE GRAFT WAS REMOVED. THE DOCTOR REPORTED THESE SYMPTOMS WENT AWAY WHEN THE GRAFT WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SEALPTFE | VASCULAR PROSTHESIS | DSY | VASCUTEK LTD. | UNK | P1747 1454 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| R |