FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG

MDR report key: 1821683 · Received August 31, 2010

Report

Report Number
3005075853-2010-04973
Event Type
Malfunction
Date Received
August 31, 2010
Date of Event
August 4, 2010
Report Date
August 6, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE (B)(4) DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH TWO WHITE RELOADS PRESENT. THE RELOADS WERE RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY PROCEDURE THE FIRST AND SECOND FIRING OF THE DEVICE WAS FINE. ON THE THIRD FIRING THE DEVICE CUT, BUT DID NOT STAPLE. NOT REPORTED HOW THE CASE WAS COMPLETED. PATIENT CONSEQUENCE NOT REPORTED. (B)(4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX ARTICNG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 (B)(4)