LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3002158293-2010-00871
- Event Type
- Malfunction
- Date Received
- August 25, 2010
- Date of Event
- August 4, 2010
- Report Date
- August 25, 2010
- Manufacturer
- ZOLL LIFECOR CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY CHARGER (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE BATTERY CHARGER WAS NOT FUNCTIONAL AND WOULD NOT CHARGE THE PT'S BATTERY PACKS. UPON EVAL, THE POWER CONNECTOR ON THE CHARGER WAS LOOSE, NOT ALLOWING THE CHARGER TO POWER ON. THE ROOT CAUSE OF THE LOOSE CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. ONCE THE POWER BRICK WAS REPLACED, THE CHARGER WAS FULLY FUNCTIONAL. NO ADVERSE EVENT RESULTED FROM THE INTERMITTENT BATTERY CHARGER. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER.
THE PT SERVICE REP (PSR) ASSISTING A (B)(6) PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT SERVICE TO REPORT AN ISSUE WITH A CHARGER. THE PSR REPORTS THE CHARGER WOULD NOT POWER ON, DESPITE PLUGGING IT INTO MULTIPLE OUTLETS. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY CHARGER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORP | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |