FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 1821417 · Received August 25, 2010

Report

Report Number
3002158293-2010-00871
Event Type
Malfunction
Date Received
August 25, 2010
Date of Event
August 4, 2010
Report Date
August 25, 2010
Manufacturer
ZOLL LIFECOR CORP
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY CHARGER (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE BATTERY CHARGER WAS NOT FUNCTIONAL AND WOULD NOT CHARGE THE PT'S BATTERY PACKS. UPON EVAL, THE POWER CONNECTOR ON THE CHARGER WAS LOOSE, NOT ALLOWING THE CHARGER TO POWER ON. THE ROOT CAUSE OF THE LOOSE CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. ONCE THE POWER BRICK WAS REPLACED, THE CHARGER WAS FULLY FUNCTIONAL. NO ADVERSE EVENT RESULTED FROM THE INTERMITTENT BATTERY CHARGER. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER.

Description of Event or Problem · 1

THE PT SERVICE REP (PSR) ASSISTING A (B)(6) PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT SERVICE TO REPORT AN ISSUE WITH A CHARGER. THE PSR REPORTS THE CHARGER WOULD NOT POWER ON, DESPITE PLUGGING IT INTO MULTIPLE OUTLETS. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY CHARGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORP WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR