LIFEVEST WCD 3000 SYSTEM
MDR report key: 1821409
·
Received August 25, 2010
Report
- Report Number
- 3002158293-2010-00882
- Event Type
- Malfunction
- Date Received
- August 25, 2010
- Date of Event
- July 20, 2010
- Report Date
- August 25, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK SN (B)(4) HAS BEEN COMPLETED. THE CAUSE OF THE DISCHARGED BATTERY PACK WAS DEFECTIVE CELLS WITHIN THE BATTERY PACK. ONE OR MORE OF THE CELLS HAD RESIDUE/CORROSION ON THEM. THE SOURCE OF THE RESIDUE/CORROSION HAS NOT BEEN POSITIVELY IDENTIFIED BUT IS LIKELY ELECTROLYTE FROM A LEAKING CELL. THE ROOT CAUSE OF THE LEAKING CELL WAS NOT POSITIVELY IDENTIFIED. THE LAST PATIENT TO USE THIS BATTERY DID NOT REPORT ANY DEFICIENCIES. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BATTERY SN (B)(4) WAS FOUND TO HAVE ONE OR MORE DEFECTIVE CELLS. THE LAST PATIENT TO USE THIS BATTERY DID NOT REPORT ANY DEFICIENCIES.