KEEPSAFE DELUXE NO POWER SWITCH
MDR report key: 1821289
·
Received August 26, 2010
Report
- Report Number
- 2020362-2010-00188
- Event Type
- Malfunction
- Date Received
- August 26, 2010
- Report Date
- June 22, 2010
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): EVAL RESULTS - INSPECTION OF THE RETURNED PRODUCT FOUND THAT THE POWER IS INTERMITTENT WHEN USING NEW BATTERIES BUT NOT WHEN USING THE AC ADAPTER. THE BATTERY DOOR IS MISSING AND THERE IS BATTERY ACID LEAKAGE AND CORROSION IN THE BATTERY COMPARTMENT. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE ALARM HAS NO POWER WHEN TEST WITH NEW BATTERIES. BATTERY CONNECTORS ARE INTACT. THE BATTERY DOOR IS MISSING FROM THE ALARM. NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM CASE. THERE WAS NO PT INJURY REPORTED. INSPECTION OF THE PRODUCT FOUND THAT THE POWER IS INTERMITTENT ON THE ALARM.