FDA Adverse Event Injury Summary report: N

POWERED WHEELCHAIR

MDR report key: 1821160 · Received August 26, 2010

Report

Report Number
1525712-2010-00124
Event Type
Injury
Date Received
August 26, 2010
Date of Event
July 1, 2010
Report Date
August 25, 2010
Manufacturer
INVACARE
Product Code
ITI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER RECEIVED COMPLAINT A USER FELL OUT OFF DUE TO A FRAME BREAKAGE. ON (B)(6) 2010, MANUFACTURER SPOKE WITH CONSUMER WHO ALLEGED THEY SOUGHT MEDICAL TREATMENT FROM THEIR HOME CARE NURSE REGARDING AN EXTRA WOUND. THE CHAIR WAS RECEIVED AND INSPECTED. THERE WERE MARKS ON THE FRAME OF THE CHAIR INDICATING THE CHAIR WAS SECURED AT SOME POINT. ADDITIONAL HARDWARE WAS OBSERVED ON THE FRONT OF THE CHAIR FURTHER SUGGESTING A TIE DOWN FUNCTION. IN CONVERSING WITH THE USER HE HAS ADVISED HE IS USING A PUBLIC TRANSPORTATION VEHICLE THAT SECURES HIS CHAIR WHILE HE IS SEATED. AS OF THIS DATE, THE DEPT OF TRANSPORTATION HAS NOT APPROVED ANY TIE-DOWN SYSTEMS FOR TRANSPORTATION OF A USER WHILE IN A WHEELCHAIR, IN A MOVING VEHICLE OF ANY TYPE. IT IS INVACARE'S POSITION THAT USERS OF WHEELCHAIRS SHOULD BE TRANSFERRED INTO APPROPRIATE SEATING IN VEHICLES FOR TRANSPORTATION AND USE BE MADE OF THE RESTRAINTS MADE AVAILABLE BY THE AUTO INDUSTRY. INVACARE CANNOT AND DOES NOT RECOMMEND ANY WHEELCHAIR TRANSPORTATION SYSTEMS. THIS INCIDENT CLEARLY IS A RESULT OF IMPROPER USE. MDR FILED BASED ON ALLEGED UNCONFIRMED MEDICAL INTERVENTION.

Description of Event or Problem · 1

THE CONSUMER ALLEGEDLY HEARD A LOUD BANG AND WHEN HE ATTEMPTED TO GET OUT OF THE CHAIR, THE WELDS ALLEGEDLY SNAPPED OFF ON THE TILT FRAME, CAUSING HIM TO FALL AND SCRAPE HIS LEGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI INVACARE 3GAR POWER CHAIR

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention