FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1821122 · Received August 28, 2010

Report

Report Number
2531779-2010-01141
Event Type
Malfunction
Date Received
August 28, 2010
Date of Event
July 29, 2010
Report Date
July 29, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AND THERE WAS VISIBLE MOISTURE INGRESS CAUSING CORROSION AROUND A BATTERY TERMINAL. THE PUMP WAS BOOTED AND EXERCISED WITH NO OVERHEATING OR POWER LOSS DUPLICATED. CURRENT DRAWS WERE CONDUCTED AND FOUND TO BE WITHIN SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP BATTERY WAS HOT TO THE TOUCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 12 YR