FDA Adverse Event Malfunction Summary report: N

45MM ARTICNG LNR CUTR/6 ROW BL

MDR report key: 1821119 · Received August 10, 2007

Report

Report Number
1527736-2007-05293
Event Type
Malfunction
Date Received
August 10, 2007
Date of Event
June 5, 2007
Report Date
June 27, 2007
Manufacturer
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS, CONCLUSIONS = DAMAGED FIRING MECHANISM. EVALUATION SUMMARY: THE ANALYSIS SHOWED THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE CARTRIDGE WAS RECEIVED PARTIALLY FIRED AND WITH THE SPRING CARTRIDGE LOCK OUT DAMAGED. THE DAMAGE TO THE SPRING CARTRIDGE LOCKOUT IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. IN ADDITION, THE INSTRUCTIONS FOR USE STATE, "THE FIRING STROKE MUST BE COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. NOTE: ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE." THE RETURNED DEVICE WAS FOUND TO BE NON FUNCTIONAL AS THE FIRING MECHANISM WAS NOTED TO BE DAMAGED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS, THE FIRING TRIGGER TEETH AND THE LEFT SHROUD PINION AXLE SUPPORT WERE FOUND DAMAGED. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTROPLASTY PROCEDURE, THE DEVICE BROKE. THERE WAS NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 45MM ARTICNG LNR CUTR/6 ROW BL GDW ETHICON ENDO SURGERY, INC. (CINCINNATI) NA D4G62R

Patients

Seq Age Sex Outcome Treatment
1