FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL

MDR report key: 1821117 · Received August 10, 2007

Report

Report Number
1527736-2007-05302
Event Type
Malfunction
Date Received
August 10, 2007
Date of Event
June 28, 2007
Report Date
July 12, 2007
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE ANALYSIS FOUND THAT INSTRUMENTS A AND B WERE RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENTS WERE CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE CONDITION OF THE JAWS. THE INSTRUMENTS LOCKED OUT AS INTENDED. INSTRUMENT C WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS. THE INSTRUMENT LOCKED OUT AS INTENDED. CORRECTIVE AND PREVENTIVE ACTIONS HAVE BEEN INITIATED TO ADDRESS THE ROOT CAUSE OF THE FINDINGS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY, DEFECTIVE CLIP. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL FZP ETHICON ENDO SURGERY, INC (CINCINNATI) NA C4FT87

Patients

Seq Age Sex Outcome Treatment
1