FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL
MDR report key: 1821115
·
Received August 9, 2007
Report
- Report Number
- 1527736-2007-05282
- Event Type
- Malfunction
- Date Received
- August 9, 2007
- Date of Event
- July 10, 2007
- Report Date
- July 11, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DROPPING OR EJECTED CLIP. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE JAWS NOT OPENING PROPERLY UPON RELEASING OF THE TRIGGER. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE FIRING ISSUE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE MISFIRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL | FZP | ETHICON ENDO SURGERY, INC (CINCINNATI) | NA | A4AT25 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |