FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL
MDR report key: 1821114
·
Received August 9, 2007
Report
- Report Number
- 1527736-2007-05278
- Event Type
- Malfunction
- Date Received
- August 9, 2007
- Report Date
- July 11, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DAMAGED JAWS. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE CONDITION OF THE JAWS. THE INSTRUMENT LOCKED OUT AS INTENDED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY THE STAPLE COULD NOT BE LOADED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.