FDA Adverse Event
Malfunction
Summary report: N
I-STAT CHEM 8+N CARTRIDGE
MDR report key: 1821079
·
Received August 27, 2010
Report
- Report Number
- 2245578-2010-00117
- Event Type
- Malfunction
- Date Received
- August 27, 2010
- Date of Event
- January 1, 2010
- Report Date
- August 27, 2010
- Manufacturer
- ABBOTT POINT OF CARE CANADA LTD.
- Product Code
- CGL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2010, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THE I-STAT CHEM 8+ CARTRIDGE YIELDED A HEMATOCRIT RESULT OF 51%PCV, ON ANALYZER (B)(4). SEPARATE SAMPLE, THE I-STAT CHEM 8+ CARTRIDGE YIELDED A HEMOCRIT RESULT OF 19%PCV, ON ANALYZER (B)(4). SEPARATE SAMPLE, TESTED USING A LABORATORY INSTRUMENT YIELDED A HEMOCRIT RESULT OF 21.5%PCV. INITIAL I-STAT HEMATOCRIT RESULT OF 51%PCV, ON ANALYZER (B)(4), DID NOT MEET THE CLINICAL PICTURE. ADD'L INFO ON PT HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT CHEM 8+N CARTRIDGE | CHEM 8+ CARTRIDGE | CGL | ABBOTT POINT OF CARE CANADA LTD. | G10146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |