FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM 8+N CARTRIDGE

MDR report key: 1821079 · Received August 27, 2010

Report

Report Number
2245578-2010-00117
Event Type
Malfunction
Date Received
August 27, 2010
Date of Event
January 1, 2010
Report Date
August 27, 2010
Manufacturer
ABBOTT POINT OF CARE CANADA LTD.
Product Code
CGL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2010, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THE I-STAT CHEM 8+ CARTRIDGE YIELDED A HEMATOCRIT RESULT OF 51%PCV, ON ANALYZER (B)(4). SEPARATE SAMPLE, THE I-STAT CHEM 8+ CARTRIDGE YIELDED A HEMOCRIT RESULT OF 19%PCV, ON ANALYZER (B)(4). SEPARATE SAMPLE, TESTED USING A LABORATORY INSTRUMENT YIELDED A HEMOCRIT RESULT OF 21.5%PCV. INITIAL I-STAT HEMATOCRIT RESULT OF 51%PCV, ON ANALYZER (B)(4), DID NOT MEET THE CLINICAL PICTURE. ADD'L INFO ON PT HAS BEEN REQUESTED.