FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1820943 · Received October 29, 2007

Report

Report Number
1527736-2007-07111
Event Type
Malfunction
Date Received
October 29, 2007
Date of Event
April 16, 2007
Report Date
April 25, 2007
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GENERATOR AND GAVE A SOLID TONE. THE HAND PIECE WAS DISASSEMBLED; A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. ADDITIONALLY, THE CONTINUITY WAS OUT OF SPECIFICATION. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE HAD AN UNK PROBLEM. THERE WERE NO DETAILS AVAILABLE ON THE PROBLEM, BUT THERE WERE SOLID TONES WHEN ACTIVATING WITH A TEST TIP DURING TROUBLESHOOTING. THERE WAS NO PT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO SURGERY, INC (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR